Recall Z-1366-2026— Medline Industries, LP
Class II · Ongoing
Class II FDA medical device recall by Medline Industries, LP, initiated 2025-12-31. It names one FDA 510(k) clearance: K242126. Product: Medline Industries IV Administration and Extension Sets: 1) DBD-IV ADMIN SET,60DRP,1NEEDLE-FREE VALV, Model Number: DYNDTB0529; 2) IV ADMIN SET, 10DRP, 1NF, 92, Model Number: DYNDTB1029; 3) DBD-IVEXTSET2.4ML2 NEEDLE-FREEVLVE,35, Model Number: DYNDTB1516; 4) DBD-IV ADMIN SET, 15DRP, 1 INJECT PORT,, Model Number: DYNDTB1530; 5) DBD-IV ADMIN SET, 15DRP, ROLLER CLAMP, 6, Model Number: DYNDTB1539; 6) DBD-IV ADMIN SET, 15DRP, 3 NF, 109, Model Number: DYNDTB1545; 7) IV ADMIN SET, 20DRP, 3 NF, 109, Model Number: DYNDTB2045; 8) DBD-IVADMINSET60DRP2NEEDLE-FREEVLVE96, Model Number: DYNDTB0512; 9) DBD-IVADMINSET,60DRP3 NEEDLE-FREE VLVE11, Model Number: DYNDTB0555; 10) IV ADMIN SET, 10DRP, NF,INJ, 96, Model Number: DYNDTB1022; 11) DBD-IV ADMIN SET, 15DRP, 2NF, 96, Model Number: DYNDTB1512; 12) DBD-IV ADMIN SET, 15DRP, 1NF, 92, Model Number: DYNDTB1529; 13) DBD-IV ADMIN SET, 15DRP, ROLLER CLAMP, 9, Model Number: DYNDTB1537; 14) DBD-IV ADMIN SET, 15DRP, SECOND, 40, Model Number: DYNDTB1540; 15) IV ADMIN SET, 20DRP, 1NF, 92, Model Number: DYNDTB2029; 16) DBD-IV EXT SET, 6.5, 1.1ML, M LUERLOCK,, Model Number: DYNDTB5077; 17) IV ADMIN SET, 10DRP, 3 NF, 109, Model Number: DYNDTB1045; 18) DBD-IVADMISET15DRP4NEEDLEFREE VLVESTPCK1, Model Number: DYNDTB1545D; 19) IV ADMIN SET15DRPFLWCTRL1NEEDLEFREE VA, Model Number: DYNDTB2524; 20) IV ADMIN SET, 10DRP, 2NF, 96, Model Number: DYNDTB1012; 21) IV ADMIN SET, 20DRP, 2NF, 96, Model Number: DYNDTB2012; 22) IV ADMIN SET, 20DRP, NF,INJ, 96, Model Number: DYNDTB2022; 23) DBD-IV EXT SET,SMBORE EXT,8,M LUERLOCK,1, Model Number: DYNDTB5081. Reason: Under circumstances involving exposure to particular chemicals and mechanical forces, the male luer connectors for the specified IV Administration and Extension sets may leak, crack, and/or break during use, which may result in a delay in patient treatment, blood loss, infection, an air embolism, and/or underdosage of medication..
- Recalling firm
- Medline Industries, LP
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2025-12-31
- Firm location
- Northfield, IL, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Medline Industries IV Administration and Extension Sets: 1) DBD-IV ADMIN SET,60DRP,1NEEDLE-FREE VALV, Model Number: DYNDTB0529; 2) IV ADMIN SET, 10DRP, 1NF, 92, Model Number: DYNDTB1029; 3) DBD-IVEXTSET2.4ML2 NEEDLE-FREEVLVE,35, Model Number: DYNDTB1516; 4) DBD-IV ADMIN SET, 15DRP, 1 INJECT PORT,, Model Number: DYNDTB1530; 5) DBD-IV ADMIN SET, 15DRP, ROLLER CLAMP, 6, Model Number: DYNDTB1539; 6) DBD-IV ADMIN SET, 15DRP, 3 NF, 109, Model Number: DYNDTB1545; 7) IV ADMIN SET, 20DRP, 3 NF, 109, Model Number: DYNDTB2045; 8) DBD-IVADMINSET60DRP2NEEDLE-FREEVLVE96, Model Number: DYNDTB0512; 9) DBD-IVADMINSET,60DRP3 NEEDLE-FREE VLVE11, Model Number: DYNDTB0555; 10) IV ADMIN SET, 10DRP, NF,INJ, 96, Model Number: DYNDTB1022; 11) DBD-IV ADMIN SET, 15DRP, 2NF, 96, Model Number: DYNDTB1512; 12) DBD-IV ADMIN SET, 15DRP, 1NF, 92, Model Number: DYNDTB1529; 13) DBD-IV ADMIN SET, 15DRP, ROLLER CLAMP, 9, Model Number: DYNDTB1537; 14) DBD-IV ADMIN SET, 15DRP, SECOND, 40, Model Number: DYNDTB1540; 15) IV ADMIN SET, 20DRP, 1NF, 92, Model Number: DYNDTB2029; 16) DBD-IV EXT SET, 6.5, 1.1ML, M LUERLOCK,, Model Number: DYNDTB5077; 17) IV ADMIN SET, 10DRP, 3 NF, 109, Model Number: DYNDTB1045; 18) DBD-IVADMISET15DRP4NEEDLEFREE VLVESTPCK1, Model Number: DYNDTB1545D; 19) IV ADMIN SET15DRPFLWCTRL1NEEDLEFREE VA, Model Number: DYNDTB2524; 20) IV ADMIN SET, 10DRP, 2NF, 96, Model Number: DYNDTB1012; 21) IV ADMIN SET, 20DRP, 2NF, 96, Model Number: DYNDTB2012; 22) IV ADMIN SET, 20DRP, NF,INJ, 96, Model Number: DYNDTB2022; 23) DBD-IV EXT SET,SMBORE EXT,8,M LUERLOCK,1, Model Number: DYNDTB5081
Reason for recall
Under circumstances involving exposure to particular chemicals and mechanical forces, the male luer connectors for the specified IV Administration and Extension sets may leak, crack, and/or break during use, which may result in a delay in patient treatment, blood loss, infection, an air embolism, and/or underdosage of medication.