Recall Z-1363-2020— Siemens Healthcare Diagnostics, Inc.

Class II · Terminated

Class II FDA medical device recall by Siemens Healthcare Diagnostics, Inc., initiated 2019-12-20, terminated 2020-12-21. FDA's recall record names no specific 510(k) clearance for this event. Product: ADVIA Chemistry Lipase, Ref. No. 01984894 B01-4840-01, Siemens Material No. 10311896, UDI 00630414482446 - Product Usage: Lipase measurements are used in the diagnosis and treatment of diseases of the pancreas, such as acute pancreatitis and obstruction of the pancreatic duct.. Reason: Firm has received customer complaints regarding failed calibrations and increased imprecision of quality control and patient samples when using this product..

Recalling firm
Siemens Healthcare Diagnostics, Inc.
Classification
Class II
Status
Terminated
Initiated
2019-12-20
Terminated
2020-12-21
Firm location
Tarrytown, NY, United States

Product description

ADVIA Chemistry Lipase, Ref. No. 01984894 B01-4840-01, Siemens Material No. 10311896, UDI 00630414482446 - Product Usage: Lipase measurements are used in the diagnosis and treatment of diseases of the pancreas, such as acute pancreatitis and obstruction of the pancreatic duct.

Reason for recall

Firm has received customer complaints regarding failed calibrations and increased imprecision of quality control and patient samples when using this product.

Data sourced from openFDA. This site is unofficial and independent of the FDA.