Recall Z-1363-2018— Orthofix, Inc
Class II · Terminated
Class II FDA medical device recall by Orthofix, Inc, initiated 2018-02-23, terminated 2019-06-18. It names one FDA 510(k) clearance: K172194. Product: ORTHOFIX, Connector System, SMALL SET SCREW, REF 79-2003. Reason: Orthofix is conducting a voluntary recall of the Connector System Large Set Screw (PN: 79-2002) with affected Lot Numbers 001 and 002, and Small Set Screw (PN: 79-2003) with affected Lot Numbers 002, 003 and 004. Through the company's manufacturing controls, size discrepancies were noted with the drive feature for both the large and small sets screws. As a result, it is possible that (1) the set screw drive feature will not accept the Set Screw Driver (PN: 79-1006) or (2) the driver may get wedged in the corners of the drive feature (false bottom) which may lead to a stripped set screw..
- Recalling firm
- Orthofix, Inc
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2018-02-23
- Terminated
- 2019-06-18
- Firm location
- Lewisville, TX, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
ORTHOFIX, Connector System, SMALL SET SCREW, REF 79-2003
Reason for recall
Orthofix is conducting a voluntary recall of the Connector System Large Set Screw (PN: 79-2002) with affected Lot Numbers 001 and 002, and Small Set Screw (PN: 79-2003) with affected Lot Numbers 002, 003 and 004. Through the company's manufacturing controls, size discrepancies were noted with the drive feature for both the large and small sets screws. As a result, it is possible that (1) the set screw drive feature will not accept the Set Screw Driver (PN: 79-1006) or (2) the driver may get wedged in the corners of the drive feature (false bottom) which may lead to a stripped set screw.
Data sourced from openFDA. This site is unofficial and independent of the FDA.