Recall Z-1359-2020— CareFusion 303, Inc.
Class I · Ongoing
Class I FDA medical device recall by CareFusion 303, Inc., initiated 2020-02-04. It names one FDA 510(k) clearance: K133532. Product: CareFusion Alaris System, Alaris PC unit models 8000 and 8015 (UDI: 10885403801518). Reason: 1)Error 255-XX-XXX results in inability to edit settings 2)Delay options programming may result in no KVO rate/therapy interruption 3)Software errors results in no low battery alarm/infusion stopping 4)Medium priority KVO/End of Infusion alarms may result in unrecognized infusion completion 5)Custom concentration data entry errors results in concentrations being lower/higher than medication orders.
- Recalling firm
- CareFusion 303, Inc.
- Classification
- Class I
- Status
- Ongoing
- Initiated
- 2020-02-04
- Firm location
- San Diego, CA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
CareFusion Alaris System, Alaris PC unit models 8000 and 8015 (UDI: 10885403801518)
Reason for recall
1)Error 255-XX-XXX results in inability to edit settings 2)Delay options programming may result in no KVO rate/therapy interruption 3)Software errors results in no low battery alarm/infusion stopping 4)Medium priority KVO/End of Infusion alarms may result in unrecognized infusion completion 5)Custom concentration data entry errors results in concentrations being lower/higher than medication orders