Recall Z-1356-2023— Philips North America Llc
Class II · Ongoing
Class II FDA medical device recall by Philips North America Llc, initiated 2021-10-21. It names 2 FDA 510(k) clearances: K173433, K163210. Product: ProxiDiagnost R90 1.0 and 1.1 multi-functional general R/F system, Model Numbers 706100 and 706110. Reason: If the system experiences an overvoltage / power surge it can cause a short circuit condition within the Main- cabinet. If this issue occurs, the user may experience an electrical fire within the cabinet, see or smell smoke emitting from the cabinet, or hear a popping sound coming from the cabinet..
- Recalling firm
- Philips North America Llc
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2021-10-21
- Firm location
- Cambridge, MA, United States
510(k) clearances named in this recall
2 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
ProxiDiagnost R90 1.0 and 1.1 multi-functional general R/F system, Model Numbers 706100 and 706110
Reason for recall
If the system experiences an overvoltage / power surge it can cause a short circuit condition within the Main- cabinet. If this issue occurs, the user may experience an electrical fire within the cabinet, see or smell smoke emitting from the cabinet, or hear a popping sound coming from the cabinet.