Recall Z-1356-2015— Medtronic Inc. Cardiac Rhythm Disease Management

Class II · Terminated

Class II FDA medical device recall by Medtronic Inc. Cardiac Rhythm Disease Management, initiated 2015-03-17, terminated 2015-10-08. It names 2 FDA 510(k) clearances: K150246, K132924. Product: Medtronic Dual Chamber Temporary External Pacemaker, Model 5392. Product Usage: The temporary pacemaker is used with a cardiac pacing lead system for temporary single or dual chamber pacing in a clinical environment by trained personnel. The temporary pacemaker can be used where short-term demand (synchronous) or asynchronous pacing is indicated for therapeutic, prophylactic, or diagnostic purposes. The temporary pacemaker must be used in an environment where the patient is monitored continuously to ensure that it is operating properly and delivering appropriate therapy to the patient. Specific indications for temporary cardiac pacing include, but are not limited to, the following: - Complete heart block - Sinus bradycardia - Sick sinus syndrome - Bradycardia with congestive heart failure - Atrial and/or ventricular arrhythmias - Cardiac arrest - Support, management, and evaluation of a patient before permanent pacemaker implantation - Support during permanent pacemaker replacement - Cardiac complications during invasive or surgical procedures - Support following cardiac surgery - Acute myocardial infarction complicated by heart block - Atrial tachyarrhythmias that require high-rate burst pacing for treatment. Reason: Possible performance issue when used with specific AA-sized (LR6) batteries. The negative terminal of a commercially available AA (LR6) battery did not maintain a sufficient connection with the battery drawer electrical contact. This issue could prevent the EPG from powering on or cause the EPG to abruptly lose primary battery power, potentially stopping delivery of pacing therapy..

Recalling firm
Medtronic Inc. Cardiac Rhythm Disease Management
Classification
Class II
Status
Terminated
Initiated
2015-03-17
Terminated
2015-10-08
Firm location
Saint Paul, MN, United States

510(k) clearances named in this recall

2 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Medtronic Dual Chamber Temporary External Pacemaker, Model 5392. Product Usage: The temporary pacemaker is used with a cardiac pacing lead system for temporary single or dual chamber pacing in a clinical environment by trained personnel. The temporary pacemaker can be used where short-term demand (synchronous) or asynchronous pacing is indicated for therapeutic, prophylactic, or diagnostic purposes. The temporary pacemaker must be used in an environment where the patient is monitored continuously to ensure that it is operating properly and delivering appropriate therapy to the patient. Specific indications for temporary cardiac pacing include, but are not limited to, the following: - Complete heart block - Sinus bradycardia - Sick sinus syndrome - Bradycardia with congestive heart failure - Atrial and/or ventricular arrhythmias - Cardiac arrest - Support, management, and evaluation of a patient before permanent pacemaker implantation - Support during permanent pacemaker replacement - Cardiac complications during invasive or surgical procedures - Support following cardiac surgery - Acute myocardial infarction complicated by heart block - Atrial tachyarrhythmias that require high-rate burst pacing for treatment

Reason for recall

Possible performance issue when used with specific AA-sized (LR6) batteries. The negative terminal of a commercially available AA (LR6) battery did not maintain a sufficient connection with the battery drawer electrical contact. This issue could prevent the EPG from powering on or cause the EPG to abruptly lose primary battery power, potentially stopping delivery of pacing therapy.