Recall Z-1355-2026— Intuitive Surgical, Inc.
Class II · Ongoing
Class II FDA medical device recall by Intuitive Surgical, Inc., initiated 2025-12-30. It names one FDA 510(k) clearance: K232610. Product: da Vinci 5 Surgeon Console Viewer Display, ASSY, DV5 CONSOLE, IS5000, Part Number: 380730-45. Reason: Robotic-assisted surgical system has Viewer Display manufactured with outdated firmware, may result in loss of one display-loss of 3D surgical field (depth perception) but 2D maintained, or loss of both displays-complete visualization loss, without fault, instruments remain active, could lead to instrument contact with tissue, resulting in injury/bleeding; may require alternate surgical modality..
- Recalling firm
- Intuitive Surgical, Inc.
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2025-12-30
- Firm location
- Sunnyvale, CA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
da Vinci 5 Surgeon Console Viewer Display, ASSY, DV5 CONSOLE, IS5000, Part Number: 380730-45
Reason for recall
Robotic-assisted surgical system has Viewer Display manufactured with outdated firmware, may result in loss of one display-loss of 3D surgical field (depth perception) but 2D maintained, or loss of both displays-complete visualization loss, without fault, instruments remain active, could lead to instrument contact with tissue, resulting in injury/bleeding; may require alternate surgical modality.