Recall Z-1355-2020— Cook Inc.
Class II · Terminated
Class II FDA medical device recall by Cook Inc., initiated 2020-01-16, terminated 2020-11-12. FDA's recall record names no specific 510(k) clearance for this event. Product: Zenith Alpha Spiral-Z Endovascular Leg, Catalog Prefix ZISL. The product is an endovascular graft.. Reason: The firm has identified factors which contributed to failures relating to thrombus formation and/or lumen occlusion reports, and is highlighting to customers key points of the IFU pertaining to the prevention of these factors. In addition, the firm has updated the Planning and Sizing Worksheet to include associated information..
- Recalling firm
- Cook Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2020-01-16
- Terminated
- 2020-11-12
- Firm location
- Bloomington, IN, United States
Product description
Zenith Alpha Spiral-Z Endovascular Leg, Catalog Prefix ZISL. The product is an endovascular graft.
Reason for recall
The firm has identified factors which contributed to failures relating to thrombus formation and/or lumen occlusion reports, and is highlighting to customers key points of the IFU pertaining to the prevention of these factors. In addition, the firm has updated the Planning and Sizing Worksheet to include associated information.