Recall Z-1352-2020— Breas Medical, Inc.
Class II · Terminated
Class II FDA medical device recall by Breas Medical, Inc., initiated 2020-01-28, terminated 2020-09-15. It names one FDA 510(k) clearance: K160481. Product: Protective Cover Vivo 55/65 (Used for Vivo 50/65), Part No. 006344. It is an accessory (i.e. protective cover) for the Vivo 50 and Vivo 65 ventilators - Product Usage: The protective cover is intended for additional protection of the Vivo 50, Vivo 55, Vivo 60 and Vivo 65 ventilators during transportation, and in hospital, institutional, and home care environments. It can be used while the Vivo ventilators is operating, for example mounted on a wheelchair, in a personal vehicle, or carried by hand. The device is not to be used with the protective cover while positioned in a warm place, such as in direct sunlight or close to a radiator. The protective cover includes the following features: transparent window for accessing front panel and buttons; port for patient circuit, cables, O2 inlet, and standby button; cooling air inlet; patient air inlet; straps for safe mounting; and cooling air outlet.. Reason: The outlet hole of the exhalation valve control pressure tube on the ventilator may become obstructed by the protective cover..
- Recalling firm
- Breas Medical, Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2020-01-28
- Terminated
- 2020-09-15
- Firm location
- North Billerica, MA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Protective Cover Vivo 55/65 (Used for Vivo 50/65), Part No. 006344. It is an accessory (i.e. protective cover) for the Vivo 50 and Vivo 65 ventilators - Product Usage: The protective cover is intended for additional protection of the Vivo 50, Vivo 55, Vivo 60 and Vivo 65 ventilators during transportation, and in hospital, institutional, and home care environments. It can be used while the Vivo ventilators is operating, for example mounted on a wheelchair, in a personal vehicle, or carried by hand. The device is not to be used with the protective cover while positioned in a warm place, such as in direct sunlight or close to a radiator. The protective cover includes the following features: transparent window for accessing front panel and buttons; port for patient circuit, cables, O2 inlet, and standby button; cooling air inlet; patient air inlet; straps for safe mounting; and cooling air outlet.
Reason for recall
The outlet hole of the exhalation valve control pressure tube on the ventilator may become obstructed by the protective cover.