Recall Z-1350-2023— Church & Dwight Inc
Class II · Ongoing
Class II FDA medical device recall by Church & Dwight Inc, initiated 2023-01-25. It names one FDA 510(k) clearance: K083716. Product: Easy Read App associated with First Response Pregnancy Test Sticks - Intended use of converting analog pregnancy test results into text pregnant or not pregnant for display on consumer devices (e.g., smart phone, iPad). Reason: Church & Dwight First Response Pregnancy Kits was marketed with the Easy Read App without a new 510k. There is also a potential for Easy Read App to be unable to reads the test results (i.e., no results reported on the APP screen) or to misread the analog results shown on the pregnancy test stick and may lead to misinterpretation of false positive or false negative tests by the user..
- Recalling firm
- Church & Dwight Inc
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2023-01-25
- Firm location
- Princeton, NJ, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Easy Read App associated with First Response Pregnancy Test Sticks - Intended use of converting analog pregnancy test results into text pregnant or not pregnant for display on consumer devices (e.g., smart phone, iPad)
Reason for recall
Church & Dwight First Response Pregnancy Kits was marketed with the Easy Read App without a new 510k. There is also a potential for Easy Read App to be unable to reads the test results (i.e., no results reported on the APP screen) or to misread the analog results shown on the pregnancy test stick and may lead to misinterpretation of false positive or false negative tests by the user.