Recall Z-1346-2013— Roche Molecular Systems, Inc.

Class II · Terminated

Class II FDA medical device recall by Roche Molecular Systems, Inc., initiated 2013-04-01, terminated 2014-08-07. FDA's recall record names no specific 510(k) clearance for this event. Product: cobas¿ 4800 system z480 instrument. Integrates fully automated total nucleic acid isolation directly from secondary sample tubes.. Reason: In rare instances, channel shifted results have been generated with cobas¿ 4800 assays run on v1.1.1 of the cobas¿ 4800 system. Raw data review indicates that, in these cases, all signals are shifted by one channel: channel 1 results report as channel 2, channel 2 results report as channel 3, channel 3 results report as channel 4, and channel 4 results report as channel 1. In the reported.

Recalling firm
Roche Molecular Systems, Inc.
Classification
Class II
Status
Terminated
Initiated
2013-04-01
Terminated
2014-08-07
Firm location
Branchburg, NJ, United States

Product description

cobas¿ 4800 system z480 instrument. Integrates fully automated total nucleic acid isolation directly from secondary sample tubes.

Reason for recall

In rare instances, channel shifted results have been generated with cobas¿ 4800 assays run on v1.1.1 of the cobas¿ 4800 system. Raw data review indicates that, in these cases, all signals are shifted by one channel: channel 1 results report as channel 2, channel 2 results report as channel 3, channel 3 results report as channel 4, and channel 4 results report as channel 1. In the reported