Recall Z-1345-2020— Philips North America, LLC
Class II · Terminated
Class II FDA medical device recall by Philips North America, LLC, initiated 2020-01-13, terminated 2023-06-22. It names one FDA 510(k) clearance: K071733. Product: Philips DigiTrak XT Holter Recorder Software version number B.1 and C.2. The DigiTrak XT Recorder is available in four different configurations. Service tag numbers for these configurations are as follows: - 453564067151 24-Hour Configuration - 453564067161 48-Hour Configuration - 453564067171 96-Hour Configuration - 453564067181 7-Day Configuration Model Number: 860322. Reason: if an AAA battery is inserted in the recorder and a user attempts to start it, or if the recorder is inserted in the docking station, the recorder will display Error: 602 and fail to function for a new patient study..
- Recalling firm
- Philips North America, LLC
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2020-01-13
- Terminated
- 2023-06-22
- Firm location
- Andover, MA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Philips DigiTrak XT Holter Recorder Software version number B.1 and C.2. The DigiTrak XT Recorder is available in four different configurations. Service tag numbers for these configurations are as follows: - 453564067151 24-Hour Configuration - 453564067161 48-Hour Configuration - 453564067171 96-Hour Configuration - 453564067181 7-Day Configuration Model Number: 860322
Reason for recall
if an AAA battery is inserted in the recorder and a user attempts to start it, or if the recorder is inserted in the docking station, the recorder will display Error: 602 and fail to function for a new patient study.