Recall Z-1345-2016— AngioDynamics Inc.

Class II · Terminated

Class II FDA medical device recall by AngioDynamics Inc., initiated 2015-11-03, terminated 2017-03-16. FDA's recall record names no specific 510(k) clearance for this event. Product: NAMIC CONVENIENCE KIT HOLYOKE HOSPITAL SPECIAL PROCEDURE KIT, REF Catalog No. 60080094, UPN Product No. H749600800941 Product Usage - This needle is intended to be placed over the guidewire to enlarge the opening in the vessel, and is then removed before sliding the catheter over the guidewire. Reason: Procedure Products initiated the recall due to the inner lumen of the needles not meeting the intended specification. Specifically, at least a portion of the affected lot is unable to accept a 0.038" guidewire as intended. In some cases the inner lumen was found to be 0.036"..

Recalling firm
AngioDynamics Inc.
Classification
Class II
Status
Terminated
Initiated
2015-11-03
Terminated
2017-03-16
Firm location
Glens Falls, NY, United States

Product description

NAMIC CONVENIENCE KIT HOLYOKE HOSPITAL SPECIAL PROCEDURE KIT, REF Catalog No. 60080094, UPN Product No. H749600800941 Product Usage - This needle is intended to be placed over the guidewire to enlarge the opening in the vessel, and is then removed before sliding the catheter over the guidewire

Reason for recall

Procedure Products initiated the recall due to the inner lumen of the needles not meeting the intended specification. Specifically, at least a portion of the affected lot is unable to accept a 0.038" guidewire as intended. In some cases the inner lumen was found to be 0.036".