Recall Z-1344-2024— Angiodynamics, Inc.
Class II · Ongoing
Class II FDA medical device recall by Angiodynamics, Inc., initiated 2024-01-30. It names one FDA 510(k) clearance: K132713. Product: MINI STICK MAX 5F X 10 CM STD .018 SS/SS NON-ECHO 2.75" PG-Used for the percutaneous introduction of a guidewire into the vascular system. Catalog Number: 45-767. Reason: Non-conformance may prevent the guidewire from passing through the introducer hub during a surgical procedure. The inability of the guidewire to pass through the introducer is due to the presence of voids in the internal lumen of the sheath hub, potential risk of the non-conformance is a delay in procedure, wherein the user may need to exchange the sheath to complete the case.
- Recalling firm
- Angiodynamics, Inc.
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2024-01-30
- Firm location
- Queensbury, NY, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
MINI STICK MAX 5F X 10 CM STD .018 SS/SS NON-ECHO 2.75" PG-Used for the percutaneous introduction of a guidewire into the vascular system. Catalog Number: 45-767
Reason for recall
Non-conformance may prevent the guidewire from passing through the introducer hub during a surgical procedure. The inability of the guidewire to pass through the introducer is due to the presence of voids in the internal lumen of the sheath hub, potential risk of the non-conformance is a delay in procedure, wherein the user may need to exchange the sheath to complete the case
Data sourced from openFDA. This site is unofficial and independent of the FDA.