Recall Z-1339-2021— Heartware, Inc.

Class I · Ongoing

Class I FDA medical device recall by Heartware, Inc., initiated 2021-02-26. FDA's recall record names no specific 510(k) clearance for this event. Product: Medtronic HVAD Monitor Data Cable, Product Number: 1575. Reason: There is the potential for the Controller power cables (AC Adapter, DC Adapter, and Battery cables), HVAD Alarm Adapter, and HVAD Monitor data cables to cause damage to the HVAD Controller ports when inserting the cables into the Controller ports, due to wear from misalignment of the cables to the port..

Recalling firm
Heartware, Inc.
Classification
Class I
Status
Ongoing
Initiated
2021-02-26
Firm location
Miami Lakes, FL, United States

Product description

Medtronic HVAD Monitor Data Cable, Product Number: 1575

Reason for recall

There is the potential for the Controller power cables (AC Adapter, DC Adapter, and Battery cables), HVAD Alarm Adapter, and HVAD Monitor data cables to cause damage to the HVAD Controller ports when inserting the cables into the Controller ports, due to wear from misalignment of the cables to the port.

Data sourced from openFDA. This site is unofficial and independent of the FDA.