Recall Z-1332-2020— Siemens Medical Solutions USA, Inc

Class II · Terminated

Class II FDA medical device recall by Siemens Medical Solutions USA, Inc, initiated 2020-01-08, terminated 2022-06-21. It names one FDA 510(k) clearance: K173639. Product: Luminos Agile Max, Model No. 10762472 - Product Usage: Luminos Agile Max is a device intended to visualize anatomical structures by converting an X-ray pattern into a visible image. The system has medical applications ranging from gastrointestinal examinations to cranial, skeletal, thoracic and lung exposures as well as examinations of the urogenital tract.. Reason: Siemens became aware of an issue of incorrect use of the bypass key overriding the built-in collision control creating risk of collision with obstacles or persons when the operator uses the override function..

Recalling firm
Siemens Medical Solutions USA, Inc
Classification
Class II
Status
Terminated
Initiated
2020-01-08
Terminated
2022-06-21
Firm location
Malvern, PA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Luminos Agile Max, Model No. 10762472 - Product Usage: Luminos Agile Max is a device intended to visualize anatomical structures by converting an X-ray pattern into a visible image. The system has medical applications ranging from gastrointestinal examinations to cranial, skeletal, thoracic and lung exposures as well as examinations of the urogenital tract.

Reason for recall

Siemens became aware of an issue of incorrect use of the bypass key overriding the built-in collision control creating risk of collision with obstacles or persons when the operator uses the override function.

Data sourced from openFDA. This site is unofficial and independent of the FDA.