Recall Z-1330-2021— Siemens Medical Solutions USA, Inc.
Class II · Ongoing
Class II FDA medical device recall by Siemens Medical Solutions USA, Inc., initiated 2021-02-04. It names one FDA 510(k) clearance: K193257. Product: Product Label - ACUSON Sequoia *** Power Input: 100-240V~, 13.0-5.4A, 50-60Hz *** (240)11148775 Model User Manual states - ACUSON Sequoia Diagnostic Ultrasound Systems. Reason: The ultrasound system averages the Mean Sac Diameter (MSD) and Gestational Sac Diameter (GSD) into the Estimated Date of Delivery (EDD) calculation. This may result in an incorrect EDD, which may influence patient management decisions regarding induction of labor and elective caesarean delivery, which may result in premature births..
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2021-02-04
- Firm location
- Issaquah, WA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Product Label - ACUSON Sequoia *** Power Input: 100-240V~, 13.0-5.4A, 50-60Hz *** (240)11148775 Model User Manual states - ACUSON Sequoia Diagnostic Ultrasound Systems
Reason for recall
The ultrasound system averages the Mean Sac Diameter (MSD) and Gestational Sac Diameter (GSD) into the Estimated Date of Delivery (EDD) calculation. This may result in an incorrect EDD, which may influence patient management decisions regarding induction of labor and elective caesarean delivery, which may result in premature births.
Data sourced from openFDA. This site is unofficial and independent of the FDA.