Recall Z-1326-2025— Beckman Coulter, Inc.

Class II · Ongoing

Class II FDA medical device recall by Beckman Coulter, Inc., initiated 2024-11-13. It names one FDA 510(k) clearance: K221225. Product: DxI 9000 Access Immunoassay Analyzer, Catalog Number C11137. Reason: The DxI 9000 Access Immunoassay Analyzer wash wheel mixer may not spin freely within the idler pulley housing. When the problem occurs, the analyzer may generate a system event for a Wash Wheel Dispense Arm Spin Mixer Error that is displayed on the navigation bar in the User Interface. The analyzer enters the red system status and will generate a SYS flag. The tests that are in progress may be cancelled and may cause a delay in reporting patient results.

Recalling firm
Beckman Coulter, Inc.
Classification
Class II
Status
Ongoing
Initiated
2024-11-13
Firm location
Chaska, MN, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

DxI 9000 Access Immunoassay Analyzer, Catalog Number C11137

Reason for recall

The DxI 9000 Access Immunoassay Analyzer wash wheel mixer may not spin freely within the idler pulley housing. When the problem occurs, the analyzer may generate a system event for a Wash Wheel Dispense Arm Spin Mixer Error that is displayed on the navigation bar in the User Interface. The analyzer enters the red system status and will generate a SYS flag. The tests that are in progress may be cancelled and may cause a delay in reporting patient results

Data sourced from openFDA. This site is unofficial and independent of the FDA.