Recall Z-1325-2019— PENTAX of America Inc

Class II · Terminated

Class II FDA medical device recall by PENTAX of America Inc, initiated 2018-12-03, terminated 2020-08-14. It names one FDA 510(k) clearance: K163684. Product: Pentax Medical C2 CryoBalloon Standard Focal Catheter, Model FG 1028, Sterile, RX. Product Usage: The C2 CryoBalloon Ablation System is intended to be used as a cryosurgical tool in the field of general surgery, specifically for endoscopic application and the ablation of dysplastic Barrett s Esophagus. The system includes the following components: (1) CryoBalloon Catheter connects to the Controller, which controls the operation of the Catheter such as diffuser (sprayer) positioning and ablation (nitrous oxide release); (2) CryoBalloon Controller contains the cartridge heater and cryogen delivery valve, which is controlled with the foot pedal; (3) Foot Pedal; and (4) CryoBalloon Cartridge containing 36 grams of nitrous oxide.. Reason: Incorrect default dose and dose increments may be transmitted to the controller if the catheter RFID tag is not correct or is not being correctly or completely readable by the controller..

Recalling firm
PENTAX of America Inc
Classification
Class II
Status
Terminated
Initiated
2018-12-03
Terminated
2020-08-14
Firm location
Redwood City, CA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Pentax Medical C2 CryoBalloon Standard Focal Catheter, Model FG 1028, Sterile, RX. Product Usage: The C2 CryoBalloon Ablation System is intended to be used as a cryosurgical tool in the field of general surgery, specifically for endoscopic application and the ablation of dysplastic Barrett s Esophagus. The system includes the following components: (1) CryoBalloon Catheter connects to the Controller, which controls the operation of the Catheter such as diffuser (sprayer) positioning and ablation (nitrous oxide release); (2) CryoBalloon Controller contains the cartridge heater and cryogen delivery valve, which is controlled with the foot pedal; (3) Foot Pedal; and (4) CryoBalloon Cartridge containing 36 grams of nitrous oxide.

Reason for recall

Incorrect default dose and dose increments may be transmitted to the controller if the catheter RFID tag is not correct or is not being correctly or completely readable by the controller.

Data sourced from openFDA. This site is unofficial and independent of the FDA.