Recall Z-1323-2023— Medtronic Sofamor Danek USA Inc
Class II · Ongoing
Class II FDA medical device recall by Medtronic Sofamor Danek USA Inc, initiated 2023-02-08. It names one FDA 510(k) clearance: K051195. Product: Medtronic Grafton Putty: a) DBM S43101 GRAFTON 0.5CC PUTTY, REF S43101; b) DBM S43102 GRAFTON 1CC PUTTY, REF S43102; c) DBM S43103 GRAFTON 2.5CC PUTTY, REF S43103; d) DBM S43105 GRAFTON 5CC PUTTY, REF S43105; e) DBM S43110 GRAFTON 10CC PUTTY, REF S43110; f) DBM T43101 0.5CC GRAFTON PUTTY, REF T43101; g) DBM T43101INT 0.5CC GRAFTON PUTTY, REF T43101INT; h) DBM T43102 1CC GRAFTON PUTTY, REF T43102; i) DBM T43102INT 1CC GRAFTON PUTTY, REF T43102INT; j) DBM T43102JPN GRAFTON PUTTY 1CC JAR, REF T43102JPN; k) DBM T43103 2.5CC GRAFTON PUTTY, REF T43103; l) DBM T43103INT 2.5CC GRAFTON PUTTY, REF T43103INT; m) DBM T43103JPN GRAFTON PUTTY 2.5CC JAR, REF T43103JPN; n) DBM T43105 5CC GRAFTON PUTTY, REF T43105; o) DBM T43105INT 5CC GRAFTON PUTTY, REF T43105INT; p) DBM T43105JPN GRAFTON PUTTY 5CC JAR, REF T43105JPN; q) DBM T43110 10CC GRAFTON PUTTY, REF T43110; r) DBM T43110INT 10CC GRAFTON PUTTY, REF T43110INT; Demineralized Bone Matrix (DBM) product. Reason: Potential for packaging non-conformances directly related to the dual-barrier, sterile pouch packaging system. Non-conforming packaging may lead to a breach in the sterile barrier..
- Recalling firm
- Medtronic Sofamor Danek USA Inc
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2023-02-08
- Firm location
- Memphis, TN, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Medtronic Grafton Putty: a) DBM S43101 GRAFTON 0.5CC PUTTY, REF S43101; b) DBM S43102 GRAFTON 1CC PUTTY, REF S43102; c) DBM S43103 GRAFTON 2.5CC PUTTY, REF S43103; d) DBM S43105 GRAFTON 5CC PUTTY, REF S43105; e) DBM S43110 GRAFTON 10CC PUTTY, REF S43110; f) DBM T43101 0.5CC GRAFTON PUTTY, REF T43101; g) DBM T43101INT 0.5CC GRAFTON PUTTY, REF T43101INT; h) DBM T43102 1CC GRAFTON PUTTY, REF T43102; i) DBM T43102INT 1CC GRAFTON PUTTY, REF T43102INT; j) DBM T43102JPN GRAFTON PUTTY 1CC JAR, REF T43102JPN; k) DBM T43103 2.5CC GRAFTON PUTTY, REF T43103; l) DBM T43103INT 2.5CC GRAFTON PUTTY, REF T43103INT; m) DBM T43103JPN GRAFTON PUTTY 2.5CC JAR, REF T43103JPN; n) DBM T43105 5CC GRAFTON PUTTY, REF T43105; o) DBM T43105INT 5CC GRAFTON PUTTY, REF T43105INT; p) DBM T43105JPN GRAFTON PUTTY 5CC JAR, REF T43105JPN; q) DBM T43110 10CC GRAFTON PUTTY, REF T43110; r) DBM T43110INT 10CC GRAFTON PUTTY, REF T43110INT; Demineralized Bone Matrix (DBM) product
Reason for recall
Potential for packaging non-conformances directly related to the dual-barrier, sterile pouch packaging system. Non-conforming packaging may lead to a breach in the sterile barrier.