Recall Z-1322-2021— EPIX THERAPEUTICS, INC

Class II · Ongoing

Class II FDA medical device recall by EPIX THERAPEUTICS, INC, initiated 2021-02-22. FDA's recall record names no specific 510(k) clearance for this event. Product: EPIX Therapeutics DiamondTemp Unidirectional Ablation Catheter, Large Curve, 7.5 French Size(2.48 mm), Connector Type: ODU 19 PIN, REF:DT200L, SterileEO, Rx Only, UDI: 00812499030013. Reason: It has been identified that the catheter size on the packaging labeling does not correspond to the widest part of the catheter. In addition the IFU contains the incorrect diameter of the catheter. This could result in blood stagnation, clots or air emboli..

Recalling firm
EPIX THERAPEUTICS, INC
Classification
Class II
Status
Ongoing
Initiated
2021-02-22
Firm location
Sunnyvale, CA, United States

Product description

EPIX Therapeutics DiamondTemp Unidirectional Ablation Catheter, Large Curve, 7.5 French Size(2.48 mm), Connector Type: ODU 19 PIN, REF:DT200L, SterileEO, Rx Only, UDI: 00812499030013

Reason for recall

It has been identified that the catheter size on the packaging labeling does not correspond to the widest part of the catheter. In addition the IFU contains the incorrect diameter of the catheter. This could result in blood stagnation, clots or air emboli.

Data sourced from openFDA. This site is unofficial and independent of the FDA.