Recall Z-1318-2024— Think Surgical, Inc.
Class II · Ongoing
Class II FDA medical device recall by Think Surgical, Inc., initiated 2024-02-12. It names one FDA 510(k) clearance: K230202. Product: Assembly, Array Guide TMINI, REF: 109101, part of the TMINI Knee Instrument Set, used with the TMINI Miniature Robotic System. Reason: There is a galling effect (cold-weld) caused by friction between the array fixation pins and the TMINI array guide holes due to insufficient clearance between the outer diameter of the fixation pins and the inner diameter of the TMINI array guide holes. As a result, the array guide has the possibility of binding to the array fixation pins during drilling..
- Recalling firm
- Think Surgical, Inc.
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2024-02-12
- Firm location
- Fremont, CA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Assembly, Array Guide TMINI, REF: 109101, part of the TMINI Knee Instrument Set, used with the TMINI Miniature Robotic System
Reason for recall
There is a galling effect (cold-weld) caused by friction between the array fixation pins and the TMINI array guide holes due to insufficient clearance between the outer diameter of the fixation pins and the inner diameter of the TMINI array guide holes. As a result, the array guide has the possibility of binding to the array fixation pins during drilling.