Recall Z-1316-2025— ZOLL Medical Corporation

Class II · Ongoing

Class II FDA medical device recall by ZOLL Medical Corporation, initiated 2025-02-13. FDA's recall record names no specific 510(k) clearance for this event. Product: Brand Name: ZOLL Product Name: Powerheart G5 AED Model/Catalog Number: (1) G5A-00A, (2) G5A-01A, (3) G5A-02A, (4) G5A-03A, (5) G5A-19A, (6) G5A-80A, (7) G5A-80A-TSO. Software Version: NA Product Description: The Powerheart G5 automated external defibrillator (AED) is designed for treating life-threatening heart beat irregularities, such as ventricular fibrillation, that cause Sudden Cardiac Arrest (SCA). There are two models available fully automatic and semi-automatic. After the defibrillations pads are applied to the patient, the fully automatic model evaluates the heart rhythm and, if a shockable rhythm is detected, delivers a shock without any rescuer assistance. The semi-automatic model evaluates the heart rhythm and requires the rescuer to press the shock button if a shockable rhythm is detected. Both models have voice and text instructions that guide the rescuer through the entire defibrillation process. Note: Not all configurations described in this document are available in all areas.. Reason: The potential for devices to fail their self-test as a result of prolonged exposure to extreme environments (temperature and/or humidity)..

Recalling firm
ZOLL Medical Corporation
Classification
Class II
Status
Ongoing
Initiated
2025-02-13
Firm location
Chelmsford, MA, United States

Product description

Brand Name: ZOLL Product Name: Powerheart G5 AED Model/Catalog Number: (1) G5A-00A, (2) G5A-01A, (3) G5A-02A, (4) G5A-03A, (5) G5A-19A, (6) G5A-80A, (7) G5A-80A-TSO. Software Version: NA Product Description: The Powerheart G5 automated external defibrillator (AED) is designed for treating life-threatening heart beat irregularities, such as ventricular fibrillation, that cause Sudden Cardiac Arrest (SCA). There are two models available fully automatic and semi-automatic. After the defibrillations pads are applied to the patient, the fully automatic model evaluates the heart rhythm and, if a shockable rhythm is detected, delivers a shock without any rescuer assistance. The semi-automatic model evaluates the heart rhythm and requires the rescuer to press the shock button if a shockable rhythm is detected. Both models have voice and text instructions that guide the rescuer through the entire defibrillation process. Note: Not all configurations described in this document are available in all areas.

Reason for recall

The potential for devices to fail their self-test as a result of prolonged exposure to extreme environments (temperature and/or humidity).

Data sourced from openFDA. This site is unofficial and independent of the FDA.