Recall Z-1315-2024— Mckesson Medical-Surgical Inc. Corporate Office

Class II · Ongoing

Class II FDA medical device recall by Mckesson Medical-Surgical Inc. Corporate Office, initiated 2024-01-09. It names one FDA 510(k) clearance: K972868. Product: Abbott Point of Care Control Control i-STAT¿ Level 3 1.7 mL. Reason: Facility Critical Temperature Excursions on 08-JAN 2024 in Walk-In refrigerator due to mechanical failures, causing temperatures to fall below the validated range of 35.6 - 46.4 Degrees Fahrenheit (2 - 8 Degrees Celsius) for up to 60 minutes which could increase the probability of an incorrect test result or a delayed test result..

Recalling firm
Mckesson Medical-Surgical Inc. Corporate Office
Classification
Class II
Status
Ongoing
Initiated
2024-01-09
Firm location
Richmond, VA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Abbott Point of Care Control Control i-STAT¿ Level 3 1.7 mL

Reason for recall

Facility Critical Temperature Excursions on 08-JAN 2024 in Walk-In refrigerator due to mechanical failures, causing temperatures to fall below the validated range of 35.6 - 46.4 Degrees Fahrenheit (2 - 8 Degrees Celsius) for up to 60 minutes which could increase the probability of an incorrect test result or a delayed test result.

Data sourced from openFDA. This site is unofficial and independent of the FDA.