Recall Z-1315-2014— Roche Diagnostics Operations, Inc.

Class II · Terminated

Class II FDA medical device recall by Roche Diagnostics Operations, Inc., initiated 2014-02-28, terminated 2018-01-10. It names one FDA 510(k) clearance: K032311. Product: Omni Micro-electrode/reference electrode for cobas b221 analyzer, Model/Catalog/Part Number: 03111873180 as a part of the following systems: 1. 03337103001, cobas b 221<1>Roche OMNI S1 system 2. 03337111001, cobas b 221<2>Roche OMNI S2 system 3. 03337154001,cobas b 221<6>Roche OMNI S6 system 4. 03337138001, cobas b 221<4>Roche OMNI S4 system Product Usage: Usage: Blood Gas and Electrolytes Analyzer c. Classification Name: Fully automated Critical Care Analyzer for the measurement of pH, Blood gases, electrolytes, Hematocrit, hemoglobin, glucose, lactate, urea/BUN, total hemoglobin, Oxygen saturation, oxy - deoxycarboxy and methemoglobin. Reason: Reference electrode used beyond the guaranteed in-use 52 week lifetime,may leak and potentially cause erroneous pH and/or sodium results. Other Ion parameters are not affected..

Recalling firm
Roche Diagnostics Operations, Inc.
Classification
Class II
Status
Terminated
Initiated
2014-02-28
Terminated
2018-01-10
Firm location
Indianapolis, IN, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Omni Micro-electrode/reference electrode for cobas b221 analyzer, Model/Catalog/Part Number: 03111873180 as a part of the following systems: 1. 03337103001, cobas b 221<1>Roche OMNI S1 system 2. 03337111001, cobas b 221<2>Roche OMNI S2 system 3. 03337154001,cobas b 221<6>Roche OMNI S6 system 4. 03337138001, cobas b 221<4>Roche OMNI S4 system Product Usage: Usage: Blood Gas and Electrolytes Analyzer c. Classification Name: Fully automated Critical Care Analyzer for the measurement of pH, Blood gases, electrolytes, Hematocrit, hemoglobin, glucose, lactate, urea/BUN, total hemoglobin, Oxygen saturation, oxy - deoxycarboxy and methemoglobin

Reason for recall

Reference electrode used beyond the guaranteed in-use 52 week lifetime,may leak and potentially cause erroneous pH and/or sodium results. Other Ion parameters are not affected.