Recall Z-1312-2026— Medline Industries, LP

Class II · Ongoing

Class II FDA medical device recall by Medline Industries, LP, initiated 2025-12-24. FDA's recall record names no specific 510(k) clearance for this event. Product: Medline 14 Fr Suction Catheter Kits SKU DYND40972 - respiratory convenience kit. Reason: A smaller open suction catheter, the 8 Fr Suction Catheter, may have been packed within the kit instead of the required 14 Fr Suction Catheter. Use of the smaller size catheter may result in hypoxia and/or respiratory failure related to partial or full airway obstruction. Airway obstruction may require immediate emergent medical intervention to prevent respiratory failure or death. Other potential outcomes include prolonged procedure, airway irritation, or prolonged dyspnea..

Recalling firm
Medline Industries, LP
Classification
Class II
Status
Ongoing
Initiated
2025-12-24
Firm location
Northfield, IL, United States

Product description

Medline 14 Fr Suction Catheter Kits SKU DYND40972 - respiratory convenience kit

Reason for recall

A smaller open suction catheter, the 8 Fr Suction Catheter, may have been packed within the kit instead of the required 14 Fr Suction Catheter. Use of the smaller size catheter may result in hypoxia and/or respiratory failure related to partial or full airway obstruction. Airway obstruction may require immediate emergent medical intervention to prevent respiratory failure or death. Other potential outcomes include prolonged procedure, airway irritation, or prolonged dyspnea.