Recall Z-1311-2018— Roche Diagnostics Corporation
Class II · Terminated
Class II FDA medical device recall by Roche Diagnostics Corporation, initiated 2017-12-22, terminated 2019-04-05. It names one FDA 510(k) clearance: K160729. Product: Elecsys BRAHMS PCT Immunoassay for the in vitro quantitative determination of PCT (procalcitonin) in human serum and plasma (K2 and K3 EDTA, Li-Heparin).. Reason: Product exhibits a decreased recovery of patient samples on the cobas e 801 module compared to the MODULAR ANALYTICS E 170, cobas e 411, 601, and 602 analyzers..
- Recalling firm
- Roche Diagnostics Corporation
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2017-12-22
- Terminated
- 2019-04-05
- Firm location
- Indianapolis, IN, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Elecsys BRAHMS PCT Immunoassay for the in vitro quantitative determination of PCT (procalcitonin) in human serum and plasma (K2 and K3 EDTA, Li-Heparin).
Reason for recall
Product exhibits a decreased recovery of patient samples on the cobas e 801 module compared to the MODULAR ANALYTICS E 170, cobas e 411, 601, and 602 analyzers.
Data sourced from openFDA. This site is unofficial and independent of the FDA.