Recall Z-1310-2014— Technidata S.A.
Class III · Terminated
Class III FDA medical device recall by Technidata S.A., initiated 2014-02-17, terminated 2014-03-31. FDA's recall record names no specific 510(k) clearance for this event. Product: TDWorkstation version 11.01.A or higher. TDWorkstation Evolution. TDWorkstation receives patient results from a connected instrument and sends them to a Laboratory Information System (LIS).. Reason: If a conflict occurs between a Demography Query and a Result Transmission by the instrument, the TDWorkstation could send an incorrect patient demography to an instrument..
- Recalling firm
- Technidata S.A.
- Classification
- Class III
- Status
- Terminated
- Initiated
- 2014-02-17
- Terminated
- 2014-03-31
- Firm location
- Montbonnot St Martin, N/A, France
Product description
TDWorkstation version 11.01.A or higher. TDWorkstation Evolution. TDWorkstation receives patient results from a connected instrument and sends them to a Laboratory Information System (LIS).
Reason for recall
If a conflict occurs between a Demography Query and a Result Transmission by the instrument, the TDWorkstation could send an incorrect patient demography to an instrument.