Recall Z-1309-2026— ETAC A/S

Class II · Ongoing

Class II FDA medical device recall by ETAC A/S, initiated 2025-11-17. FDA's recall record names no specific 510(k) clearance for this event. Product: Brand Name: Molift Product Name: Molift Mover 300 Model/Catalog Numbers: (1) M1501, (2) M1502, (3) M1504, (4) M15050, (5) M15151, (6) M15001, (7) M15002; Product Description: The mobile hoist is an assistive device intended for alleviation of, or compensation for, a functional, impairment due to an injury or disability. The device is designed for an individual lacking the ability to stand up and transfer themselves to / from a bed, the floor, a wheelchair, a chair, a toilet or similar due to reduced mobility or physical strength. Component: The deformation occurs between the bolt and the fork in the joint connecting the lifting bar and the arm on the Molift Mover.. Reason: A customer experienced that during assembly and load testing of the device, the bolt for mounting the lifting bar broke. Upon further investigation of this customer complaint, it became clear that the defect occurs as a result of deformation of the materials involved between the lifting bar and the arm of the device. Such deformation can lead to limited mobility in the joint, which prevents proper functionality of the equipment..

Recalling firm
ETAC A/S
Classification
Class II
Status
Ongoing
Initiated
2025-11-17
Firm location
Gedved, N/A, Denmark

Product description

Brand Name: Molift Product Name: Molift Mover 300 Model/Catalog Numbers: (1) M1501, (2) M1502, (3) M1504, (4) M15050, (5) M15151, (6) M15001, (7) M15002; Product Description: The mobile hoist is an assistive device intended for alleviation of, or compensation for, a functional, impairment due to an injury or disability. The device is designed for an individual lacking the ability to stand up and transfer themselves to / from a bed, the floor, a wheelchair, a chair, a toilet or similar due to reduced mobility or physical strength. Component: The deformation occurs between the bolt and the fork in the joint connecting the lifting bar and the arm on the Molift Mover.

Reason for recall

A customer experienced that during assembly and load testing of the device, the bolt for mounting the lifting bar broke. Upon further investigation of this customer complaint, it became clear that the defect occurs as a result of deformation of the materials involved between the lifting bar and the arm of the device. Such deformation can lead to limited mobility in the joint, which prevents proper functionality of the equipment.

Data sourced from openFDA. This site is unofficial and independent of the FDA.