Recall Z-1308-2021— Biomerieux Inc

Class II · Terminated

Class II FDA medical device recall by Biomerieux Inc, initiated 2021-01-28, terminated 2023-01-12. It names one FDA 510(k) clearance: K050002. Product: VITEK 2 Compact (version numbers 8.01, 8.02 (8.0x) and 9.01, 9.02 (9.0x) with the external communication configured to HL7 (system, test, automated, antimicrobial susceptibility, short incubation). VITEK 2 is an automated system consisting of instruments, software and reagent cards designed for the identification and antimicrobial susceptibility testing of bacteria and yeast. The VITEK 2 utilizes growth-based biochemical patterns to determine identification. The VITEK 2 provides minimal inhibitory concentration (MIC) results for most organism/drug combinations as well as a category interpretation (S, I, or R) consistent with accepted criteria.. Reason: bioM¿rieux has identified a potential safety risk worst case of a false susceptible erroneous test result associated with this event. The problem is when HL7 Connection is used, the results in the VITEK¿ 2 do not match the results sent to the LIS and the Laboratory Technician would need to change results in the LIS to match those in the VITEK¿ 2. If the isolate is sent more than once to LIS, a software defect prevents the system from sending the expertised interpretation results and the Therapeutic Corrections (TC) is not sent to Laboratory Information Systems (LIS)..

Recalling firm
Biomerieux Inc
Classification
Class II
Status
Terminated
Initiated
2021-01-28
Terminated
2023-01-12
Firm location
Hazelwood, MO, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

VITEK 2 Compact (version numbers 8.01, 8.02 (8.0x) and 9.01, 9.02 (9.0x) with the external communication configured to HL7 (system, test, automated, antimicrobial susceptibility, short incubation). VITEK 2 is an automated system consisting of instruments, software and reagent cards designed for the identification and antimicrobial susceptibility testing of bacteria and yeast. The VITEK 2 utilizes growth-based biochemical patterns to determine identification. The VITEK 2 provides minimal inhibitory concentration (MIC) results for most organism/drug combinations as well as a category interpretation (S, I, or R) consistent with accepted criteria.

Reason for recall

bioM¿rieux has identified a potential safety risk worst case of a false susceptible erroneous test result associated with this event. The problem is when HL7 Connection is used, the results in the VITEK¿ 2 do not match the results sent to the LIS and the Laboratory Technician would need to change results in the LIS to match those in the VITEK¿ 2. If the isolate is sent more than once to LIS, a software defect prevents the system from sending the expertised interpretation results and the Therapeutic Corrections (TC) is not sent to Laboratory Information Systems (LIS).

Data sourced from openFDA. This site is unofficial and independent of the FDA.