Recall Z-1301-2018— The Binding Site Group, Ltd.
Class II · Terminated
Class II FDA medical device recall by The Binding Site Group, Ltd., initiated 2018-03-09, terminated 2020-06-22. It names one FDA 510(k) clearance: K122066. Product: Binding Site Optilite Beta-2-Microglobulin Urine Kit, Ref LK043.L.OPT.A, IVD. The firm name on the label is The Binding Site Group Ltd., Birmingham, U.K. The Optilite Beta-2-Microglobulin (¿2M) Urine kit is intended for the quantitative in-vitro measurement of beta-2-microglobulin in urine using the Binding Site Optilite analyser to aid in the diagnosis of active rheumatoid arthritis and kidney disease. This test should be used in conjunction with other laboratory and clinical findings.. Reason: The product is not meeting the antigen excess protection claim of up to 760mg/L, as stated in the Instructions for Use..
- Recalling firm
- The Binding Site Group, Ltd.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2018-03-09
- Terminated
- 2020-06-22
- Firm location
- Birmingham, N/A, United Kingdom
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Binding Site Optilite Beta-2-Microglobulin Urine Kit, Ref LK043.L.OPT.A, IVD. The firm name on the label is The Binding Site Group Ltd., Birmingham, U.K. The Optilite Beta-2-Microglobulin (¿2M) Urine kit is intended for the quantitative in-vitro measurement of beta-2-microglobulin in urine using the Binding Site Optilite analyser to aid in the diagnosis of active rheumatoid arthritis and kidney disease. This test should be used in conjunction with other laboratory and clinical findings.
Reason for recall
The product is not meeting the antigen excess protection claim of up to 760mg/L, as stated in the Instructions for Use.