Recall Z-1300-2025— GE Medical Systems, LLC

Class II · Ongoing

Class II FDA medical device recall by GE Medical Systems, LLC, initiated 2025-02-18. It names 2 FDA 510(k) clearances: K213715, K191777. Product: GE Healthcare Revolution Apex Plus, System, X-ray, Tomography, Computed. Reason: GE HealthCare has become aware of a potential issue on certain Revolution CT and Revolution Apex series CT systems (please see affected product details and the Appendix) that can result in coolant fluid (glycol) leaking onto the floor near the base of the gantry. The fluid is visible as it has a blue color. It is not corrosive and is not at a harmful temperature. However, there is a potential for injury if the fluid is not noticed and a person slips and/or falls..

Recalling firm
GE Medical Systems, LLC
Classification
Class II
Status
Ongoing
Initiated
2025-02-18
Firm location
Waukesha, WI, United States

510(k) clearances named in this recall

2 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

GE Healthcare Revolution Apex Plus, System, X-ray, Tomography, Computed

Reason for recall

GE HealthCare has become aware of a potential issue on certain Revolution CT and Revolution Apex series CT systems (please see affected product details and the Appendix) that can result in coolant fluid (glycol) leaking onto the floor near the base of the gantry. The fluid is visible as it has a blue color. It is not corrosive and is not at a harmful temperature. However, there is a potential for injury if the fluid is not noticed and a person slips and/or falls.

Data sourced from openFDA. This site is unofficial and independent of the FDA.