Recall Z-1299-2026— FUJIFILM Healthcare Americas Corporation

Class II · Ongoing

Class II FDA medical device recall by FUJIFILM Healthcare Americas Corporation, initiated 2026-01-09. It names one FDA 510(k) clearance: K152294. Product: FDR Visionary Suite; Model Number: CH-200; Version: (1) 566-16130-23, (2) 566-16130-31, (3). Reason: It was found that the contacts of the electromagnetic contactors used inside the cabinet of the X-ray high-voltage generator were welded, and the current might continue to be supplied to the power supply circuit even if the power of the device is cut off. As a result, it cannot be denied that the circuit in question overheats and damages the surrounding parts. It may also cause smoke..

Recalling firm
FUJIFILM Healthcare Americas Corporation
Classification
Class II
Status
Ongoing
Initiated
2026-01-09
Firm location
Lexington, MA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

FDR Visionary Suite; Model Number: CH-200; Version: (1) 566-16130-23, (2) 566-16130-31, (3)

Reason for recall

It was found that the contacts of the electromagnetic contactors used inside the cabinet of the X-ray high-voltage generator were welded, and the current might continue to be supplied to the power supply circuit even if the power of the device is cut off. As a result, it cannot be denied that the circuit in question overheats and damages the surrounding parts. It may also cause smoke.

Data sourced from openFDA. This site is unofficial and independent of the FDA.