Recall Z-1298-2019— Siemens Medical Solutions USA, Inc
Class II · Terminated
Class II FDA medical device recall by Siemens Medical Solutions USA, Inc, initiated 2019-03-18, terminated 2020-04-16. It names one FDA 510(k) clearance: K132094. Product: Cios Alpha, Model Number 10308191 Mobile X-ray system - Product Usage: The Cios Alpha is a mobile X-Ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications. Clinical applications may include but are not limited to interventional fluoroscopic, gastro-intestinal, endoscopic, urologic, pain management, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedures. The patient population may include pediatric patients.. Reason: Continuous operation at very high tube output may result in increased wear of the radiation emitting components. This may occur when tube voltages of 125 kV are used for a long period, e.g. 30 minutes or more of continuous use. As a result, risk of premature failure of the Monoblock and loss of imaging X-rays..
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2019-03-18
- Terminated
- 2020-04-16
- Firm location
- Malvern, PA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Cios Alpha, Model Number 10308191 Mobile X-ray system - Product Usage: The Cios Alpha is a mobile X-Ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications. Clinical applications may include but are not limited to interventional fluoroscopic, gastro-intestinal, endoscopic, urologic, pain management, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedures. The patient population may include pediatric patients.
Reason for recall
Continuous operation at very high tube output may result in increased wear of the radiation emitting components. This may occur when tube voltages of 125 kV are used for a long period, e.g. 30 minutes or more of continuous use. As a result, risk of premature failure of the Monoblock and loss of imaging X-rays.
Data sourced from openFDA. This site is unofficial and independent of the FDA.