Recall Z-1288-2023— Philips Respironics, Inc.

Class I · Ongoing

Class I FDA medical device recall by Philips Respironics, Inc., initiated 2023-03-01. It names one FDA 510(k) clearance: K181166. Product: Trilogy EV300 Portable Electric Ventilator, Model Numbers DS2200X11B, IN2200X15B, FX2200X15B, CA2200X12B, CN2200X17B, BR2200X18B, KR2200X15B, IA2200X15B, FR2200X14B, ND2200X15B, IT2200X21B, ES2200X15B, DE2200X13B, BL2200X15B, GB2200X15B, EU2200X15B, EU2200X19, EE2200X15B, TR2200X15B, RU2200X15B. Reason: The accuracy of delivered oxygen may deviate below the required tolerance of 5% from setpoint when providing high concentration oxygen therapy. Additionally, if equipped, the internal FiO2 sensor may indicate a value higher than the device is actually delivering. This may lead to under delivery of oxygen..

Recalling firm
Philips Respironics, Inc.
Classification
Class I
Status
Ongoing
Initiated
2023-03-01
Firm location
Murrysville, PA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Trilogy EV300 Portable Electric Ventilator, Model Numbers DS2200X11B, IN2200X15B, FX2200X15B, CA2200X12B, CN2200X17B, BR2200X18B, KR2200X15B, IA2200X15B, FR2200X14B, ND2200X15B, IT2200X21B, ES2200X15B, DE2200X13B, BL2200X15B, GB2200X15B, EU2200X15B, EU2200X19, EE2200X15B, TR2200X15B, RU2200X15B

Reason for recall

The accuracy of delivered oxygen may deviate below the required tolerance of 5% from setpoint when providing high concentration oxygen therapy. Additionally, if equipped, the internal FiO2 sensor may indicate a value higher than the device is actually delivering. This may lead to under delivery of oxygen.

Data sourced from openFDA. This site is unofficial and independent of the FDA.