Recall Z-1283-2023— DeRoyal Industries Inc
Class II · Ongoing
Class II FDA medical device recall by DeRoyal Industries Inc, initiated 2023-02-09. FDA's recall record names no specific 510(k) clearance for this event. Product: DeRoyal SafeLiner Suction Canister, 1000cc, REF PHESL-1000B. Reason: The reason for the voluntary recall is due to shrinkage of the canister lid. This shrinkage causes the lid not to contact the outer shell to create a seal and thus, it does not create and hold vacuum..
- Recalling firm
- DeRoyal Industries Inc
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2023-02-09
- Firm location
- Powell, TN, United States
Product description
DeRoyal SafeLiner Suction Canister, 1000cc, REF PHESL-1000B
Reason for recall
The reason for the voluntary recall is due to shrinkage of the canister lid. This shrinkage causes the lid not to contact the outer shell to create a seal and thus, it does not create and hold vacuum.
Data sourced from openFDA. This site is unofficial and independent of the FDA.