Recall Z-1282-2024— Fresenius Kabi USA, LLC
Class II · Completed
Class II FDA medical device recall by Fresenius Kabi USA, LLC, initiated 2022-09-16. It names one FDA 510(k) clearance: K183311. Product: LVP software of the Ivenix Infusion System (IIS), LVP-SW-0004. Reason: Retroactively reported correction from 2022: A software defect may cause an incorrect (Fail-Stop) Alarm when an administration set is loaded with an upstream occlusion present, resulting in a non-recoverable (Fail-Stop) Pump Problem alarm. This may lead to delay of therapy, and patient may experience from no to minimal to serious harm. This issue was resolved in all fielded product in software version 5.2.1, which was installed in September 2022..
- Recalling firm
- Fresenius Kabi USA, LLC
- Classification
- Class II
- Status
- Completed
- Initiated
- 2022-09-16
- Firm location
- North Andover, MA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
LVP software of the Ivenix Infusion System (IIS), LVP-SW-0004
Reason for recall
Retroactively reported correction from 2022: A software defect may cause an incorrect (Fail-Stop) Alarm when an administration set is loaded with an upstream occlusion present, resulting in a non-recoverable (Fail-Stop) Pump Problem alarm. This may lead to delay of therapy, and patient may experience from no to minimal to serious harm. This issue was resolved in all fielded product in software version 5.2.1, which was installed in September 2022.
Data sourced from openFDA. This site is unofficial and independent of the FDA.