Recall Z-1280-2024— Siemens Healthcare Diagnostics, Inc.
Class II · Ongoing
Class II FDA medical device recall by Siemens Healthcare Diagnostics, Inc., initiated 2024-01-24. It names one FDA 510(k) clearance: K183088. Product: Atellica IM Erythropoietin (EPO) Assay (100 Test). Reason: Siemens Healthcare Diagnostics has confirmed an average negative bias of -35% compared to the 3rd World Health Organization (WHO) International Standard (NIBSC code: 11/170) that is proportional across the assay measuring interval. The reference interval as claimed in the Instructions for Use (IFU) is not achieved with affected lots..
- Recalling firm
- Siemens Healthcare Diagnostics, Inc.
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2024-01-24
- Firm location
- Tarrytown, NY, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Atellica IM Erythropoietin (EPO) Assay (100 Test)
Reason for recall
Siemens Healthcare Diagnostics has confirmed an average negative bias of -35% compared to the 3rd World Health Organization (WHO) International Standard (NIBSC code: 11/170) that is proportional across the assay measuring interval. The reference interval as claimed in the Instructions for Use (IFU) is not achieved with affected lots.
Data sourced from openFDA. This site is unofficial and independent of the FDA.