Recall Z-1280-2021— Baxter Healthcare Corporation

Class II · Terminated

Class II FDA medical device recall by Baxter Healthcare Corporation, initiated 2021-01-28, terminated 2024-06-13. FDA's recall record names no specific 510(k) clearance for this event. Product: Preventative Maintenance Kits G5010007) and spare part kits (G5064801 and G5006203) - Product Usage: For Prismaflex Control Units, used for Continuous Renal Replacement Therapy (CRRT).. Reason: Variability in the performance of the tubing in the Prismaflex Control Unit ARPS (Automatic Repositioning System) Pump Assembly, may lead to alarm situations during or after a system self-test..

Recalling firm
Baxter Healthcare Corporation
Classification
Class II
Status
Terminated
Initiated
2021-01-28
Terminated
2024-06-13
Firm location
Deerfield, IL, United States

Product description

Preventative Maintenance Kits G5010007) and spare part kits (G5064801 and G5006203) - Product Usage: For Prismaflex Control Units, used for Continuous Renal Replacement Therapy (CRRT).

Reason for recall

Variability in the performance of the tubing in the Prismaflex Control Unit ARPS (Automatic Repositioning System) Pump Assembly, may lead to alarm situations during or after a system self-test.