Recall Z-1277-2023— Datascope Corp.

Class II · Ongoing

Class II FDA medical device recall by Datascope Corp., initiated 2023-02-28. It names one FDA 510(k) clearance: K112372. Product: Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP), Model Numbers Model Numbers 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85. Reason: There is an issue related to the Safety Disk not meeting a performance requirement that may impact Intra-Aortic Balloon catheter (IAB) displacement volume. This potential failure of an impacted Safety Disk may lead to slightly reduced IAB inflation during therapy or a reduction in the degree of augmentation provided to the patient during therapy..

Recalling firm
Datascope Corp.
Classification
Class II
Status
Ongoing
Initiated
2023-02-28
Firm location
Mahwah, NJ, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP), Model Numbers Model Numbers 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85

Reason for recall

There is an issue related to the Safety Disk not meeting a performance requirement that may impact Intra-Aortic Balloon catheter (IAB) displacement volume. This potential failure of an impacted Safety Disk may lead to slightly reduced IAB inflation during therapy or a reduction in the degree of augmentation provided to the patient during therapy.

Data sourced from openFDA. This site is unofficial and independent of the FDA.