Recall Z-1277-2018— Halyard Health, Inc

Class II · Terminated

Class II FDA medical device recall by Halyard Health, Inc, initiated 2017-12-13, terminated 2021-11-04. It names one FDA 510(k) clearance: K083210. Product: CORTRAK* 2 Nasogastric/Nasointestinal (NG/NI) Feeding Tubes with Electromagnetic Transmitting Stylet and CORFLO* Enteral Feeding tube, ANTI-IV* Connector, 10 Fr, 140 cm The device is used in patients that require intermittent or continuous tube feedings via the nasogastric or nasoenteric route.. Reason: The transmitting stylet within CORTRAK* 2 nasogastric/Nasointestinal (NG/NI) Feeding Tubes with ANTI-IV* connector (Product Code 20-9551Trak2A) may be visible within the window of the exit port of the feeding tube. Based on specifications, the transmitting stylet in the NG tube should not be visible within the port window..

Recalling firm
Halyard Health, Inc
Classification
Class II
Status
Terminated
Initiated
2017-12-13
Terminated
2021-11-04
Firm location
Alpharetta, GA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

CORTRAK* 2 Nasogastric/Nasointestinal (NG/NI) Feeding Tubes with Electromagnetic Transmitting Stylet and CORFLO* Enteral Feeding tube, ANTI-IV* Connector, 10 Fr, 140 cm The device is used in patients that require intermittent or continuous tube feedings via the nasogastric or nasoenteric route.

Reason for recall

The transmitting stylet within CORTRAK* 2 nasogastric/Nasointestinal (NG/NI) Feeding Tubes with ANTI-IV* connector (Product Code 20-9551Trak2A) may be visible within the window of the exit port of the feeding tube. Based on specifications, the transmitting stylet in the NG tube should not be visible within the port window.