Recall Z-1276-2016— Roche Diagnostics Operations, Inc.
Class II · Terminated
Class II FDA medical device recall by Roche Diagnostics Operations, Inc., initiated 2016-03-07, terminated 2018-01-10. FDA's recall record names no specific 510(k) clearance for this event. Product: Estradiol III; Elecsys and cobas e analyzers; REF 06656021 190; M 6.5 mL, R1 9 mL, R2 9 mL, For USA: CONTENT M Streptavidin-coated microparticles 0.72 mg/ml; R1 Two biotinylated monoclonal anti-estradiol antibodies (rabbit) 2.5 ng/m and 4.5 ng/mL Mesterolone 130 ng/mL; R2 Estradiol derivative, labeled with ruthenium complex 4.5 ng/mL; Rx only Product Usage: Immunoassay for the in vitro quantitative determination of estradiol in human serum and plasma. The electrochemiluminescence immunoassay ECLIA is intended for use on Elecsys and cobas e immunoassay analyzers.. Reason: Due to the risk of a recently identified cross reactivity, the Elecsys Estradiol assay should not be used when monitoring patient being treated with fulvestrant. Customers should be informed that fulvestrant will increase the apparent concentration of estradiol in women being treated with this drug. If treatment with fulvestrant has been altered or discontinued as a result of falsely elevated e.
- Recalling firm
- Roche Diagnostics Operations, Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2016-03-07
- Terminated
- 2018-01-10
- Firm location
- Indianapolis, IN, United States
Product description
Estradiol III; Elecsys and cobas e analyzers; REF 06656021 190; M 6.5 mL, R1 9 mL, R2 9 mL, For USA: CONTENT M Streptavidin-coated microparticles 0.72 mg/ml; R1 Two biotinylated monoclonal anti-estradiol antibodies (rabbit) 2.5 ng/m and 4.5 ng/mL Mesterolone 130 ng/mL; R2 Estradiol derivative, labeled with ruthenium complex 4.5 ng/mL; Rx only Product Usage: Immunoassay for the in vitro quantitative determination of estradiol in human serum and plasma. The electrochemiluminescence immunoassay ECLIA is intended for use on Elecsys and cobas e immunoassay analyzers.
Reason for recall
Due to the risk of a recently identified cross reactivity, the Elecsys Estradiol assay should not be used when monitoring patient being treated with fulvestrant. Customers should be informed that fulvestrant will increase the apparent concentration of estradiol in women being treated with this drug. If treatment with fulvestrant has been altered or discontinued as a result of falsely elevated e
Data sourced from openFDA. This site is unofficial and independent of the FDA.