Recall Z-1275-2022— Olympus Corporation of the Americas

Class II · Ongoing

Class II FDA medical device recall by Olympus Corporation of the Americas, initiated 2022-04-22. FDA's recall record names no specific 510(k) clearance for this event. Product: EVIS EXERA II Gastrointestinal Videoscope-indicated for use within the upper digestive tract (including the esophagus, stomach, and duodenum). Model: GIF-HQ190. Reason: Part (A-rubber) intended for a different model of gastrointestinal videoscope may have been used when replacing part on GIF-HQ190 during the last repair at Olympus. An Incorrect A-rubber fits tighter on the GIF-HQ190 which would not meet the intended design specifications.

Recalling firm
Olympus Corporation of the Americas
Classification
Class II
Status
Ongoing
Initiated
2022-04-22
Firm location
Center Valley, PA, United States

Product description

EVIS EXERA II Gastrointestinal Videoscope-indicated for use within the upper digestive tract (including the esophagus, stomach, and duodenum). Model: GIF-HQ190

Reason for recall

Part (A-rubber) intended for a different model of gastrointestinal videoscope may have been used when replacing part on GIF-HQ190 during the last repair at Olympus. An Incorrect A-rubber fits tighter on the GIF-HQ190 which would not meet the intended design specifications

Data sourced from openFDA. This site is unofficial and independent of the FDA.