Recall Z-1272-2022— Smiths Medical ASD Inc.
Class I · Ongoing
Class I FDA medical device recall by Smiths Medical ASD Inc., initiated 2022-04-18. It names one FDA 510(k) clearance: K040899. Product: Medfusion Syringe Pump Models: 3500, 3500-0600-00, 3500-0600-01, 3500-0600-249, 3500-0600-50, 3500-0600-51, 3500-0600-82, 3500-306, 3500-402, 3500-414, 3500-415, 3500-500, 3500BC, 3500E, 3500G, 3500SD, 3500SD-500, 3500VX, 3500VX-306, 3500VX-414, 3500VX-415, 3500VX-500, 3500ZE, Software Versions 6.0.0, 6.0.1. Reason: Multiple issues with the potential for interruption of therapy or over-infusion: 1. Primary Audible Alarm (PAA), 2. Unanticipated Depleted Battery Alarms, 3. Time Base Alarm, 4. Intermittent Volume Over Time (IVOT) - Infusion Continues after System Failure, 5. Clearing of Program Volume Delivered (PVD), 6. False Alarm for Rate Below Recommended Minimum for Syringe Size, 7. Incorrect Bolus or Loading Dose Time Display, 8. Domain Name Server (DNS) Port 1001.
- Recalling firm
- Smiths Medical ASD Inc.
- Classification
- Class I
- Status
- Ongoing
- Initiated
- 2022-04-18
- Firm location
- Minneapolis, MN, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Medfusion Syringe Pump Models: 3500, 3500-0600-00, 3500-0600-01, 3500-0600-249, 3500-0600-50, 3500-0600-51, 3500-0600-82, 3500-306, 3500-402, 3500-414, 3500-415, 3500-500, 3500BC, 3500E, 3500G, 3500SD, 3500SD-500, 3500VX, 3500VX-306, 3500VX-414, 3500VX-415, 3500VX-500, 3500ZE, Software Versions 6.0.0, 6.0.1
Reason for recall
Multiple issues with the potential for interruption of therapy or over-infusion: 1. Primary Audible Alarm (PAA), 2. Unanticipated Depleted Battery Alarms, 3. Time Base Alarm, 4. Intermittent Volume Over Time (IVOT) - Infusion Continues after System Failure, 5. Clearing of Program Volume Delivered (PVD), 6. False Alarm for Rate Below Recommended Minimum for Syringe Size, 7. Incorrect Bolus or Loading Dose Time Display, 8. Domain Name Server (DNS) Port 1001
Data sourced from openFDA. This site is unofficial and independent of the FDA.