Recall Z-1272-2016— Focus Diagnostics Inc
Class I · Terminated
Class I FDA medical device recall by Focus Diagnostics Inc, initiated 2016-02-10, terminated 2016-06-16. It names 2 FDA 510(k) clearances: K150962, K141458. Product: Simplexa HSV 1&2 Direct (MOL2150) with Direct Amplification Disc (MOL1451, MOL1452, MOL1455) Model Number: MOL2150, MOL1451, MOL1452, MOL1455. Reason: Focus Diagnostics is recalling the Direct Amplification Discs (DAD) supplied with the Simplexa HSV 1 & 2 Direct (MOL2150), Simplexa Flu A/B & RSV Direct (MOL2650) and Simplexa Group A Strep Direct kits (MOL2850) because it may result in a false negative result, a false positive result or Error code (s) 500 or 505, and it may result in higher invalid rates..
- Recalling firm
- Focus Diagnostics Inc
- Classification
- Class I
- Status
- Terminated
- Initiated
- 2016-02-10
- Terminated
- 2016-06-16
- Firm location
- Cypress, CA, United States
510(k) clearances named in this recall
2 on FDA's record- K150962 — Simplexa HSV 1 & 2 Direct, Simplexa HSV 1 & 2 Positive Control Pack
- K141458 — SIMPLEXA HSV 1&2 DIRECT/POSITIVE CONTROL PACK
K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Simplexa HSV 1&2 Direct (MOL2150) with Direct Amplification Disc (MOL1451, MOL1452, MOL1455) Model Number: MOL2150, MOL1451, MOL1452, MOL1455
Reason for recall
Focus Diagnostics is recalling the Direct Amplification Discs (DAD) supplied with the Simplexa HSV 1 & 2 Direct (MOL2150), Simplexa Flu A/B & RSV Direct (MOL2650) and Simplexa Group A Strep Direct kits (MOL2850) because it may result in a false negative result, a false positive result or Error code (s) 500 or 505, and it may result in higher invalid rates.