Recall Z-1270-2018— Roche Diagnostics Corporation

Class II · Terminated

Class II FDA medical device recall by Roche Diagnostics Corporation, initiated 2017-12-18, terminated 2019-04-09. FDA's recall record names no specific 510(k) clearance for this event. Product: cobas e 411 Immunoassay Analyzer (Discrete Photometric Analyzer for Clinical use), Catalog Numbers: 04775201001 (rack system), 04775279001 (disk system), 04775201973 (rack system) & 04775279973 (disk system).. Reason: Complaints have been received concerning questionable results on analyzers due to customers utilizing 13 mm sample tubes without tube adapters..

Recalling firm
Roche Diagnostics Corporation
Classification
Class II
Status
Terminated
Initiated
2017-12-18
Terminated
2019-04-09
Firm location
Indianapolis, IN, United States

Product description

cobas e 411 Immunoassay Analyzer (Discrete Photometric Analyzer for Clinical use), Catalog Numbers: 04775201001 (rack system), 04775279001 (disk system), 04775201973 (rack system) & 04775279973 (disk system).

Reason for recall

Complaints have been received concerning questionable results on analyzers due to customers utilizing 13 mm sample tubes without tube adapters.