Recall Z-1270-2014— Lin-Zhi International Inc
Class III · Terminated
Class III FDA medical device recall by Lin-Zhi International Inc, initiated 2014-02-18, terminated 2014-04-25. It names one FDA 510(k) clearance: K081008. Product: Single Analyte DAU Calibrator (5 mL) Norbuprenorphine: Cutoff Calibrator; IVD - for in vitro diagnostic use only; Lin-Zhi International Sunnyvale, CA The Norbuprenorphine Drug of Abuse (DAU) cutoff Calibrator is used for qualitative and semi-qualitative calibration of the Lin-Zhi Internal, Inc. Buprenorphine (BUP) enzyme Immunoassay (REF # A53684) on the Synchron Clinical Analyzer. Reason: Customer reported that the Synchron NBUP cutoff calibrator (10 ng/mL) could not be qualified through quality control testing. The cutoff calibrator was recovering at a concentration close to the NBUP Control Level 1 (7 ng/mL)..
- Recalling firm
- Lin-Zhi International Inc
- Classification
- Class III
- Status
- Terminated
- Initiated
- 2014-02-18
- Terminated
- 2014-04-25
- Firm location
- Sunnyvale, CA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Single Analyte DAU Calibrator (5 mL) Norbuprenorphine: Cutoff Calibrator; IVD - for in vitro diagnostic use only; Lin-Zhi International Sunnyvale, CA The Norbuprenorphine Drug of Abuse (DAU) cutoff Calibrator is used for qualitative and semi-qualitative calibration of the Lin-Zhi Internal, Inc. Buprenorphine (BUP) enzyme Immunoassay (REF # A53684) on the Synchron Clinical Analyzer
Reason for recall
Customer reported that the Synchron NBUP cutoff calibrator (10 ng/mL) could not be qualified through quality control testing. The cutoff calibrator was recovering at a concentration close to the NBUP Control Level 1 (7 ng/mL).
Data sourced from openFDA. This site is unofficial and independent of the FDA.