Recall Z-1269-2018— Roche Diagnostics Corporation
Class II · Terminated
Class II FDA medical device recall by Roche Diagnostics Corporation, initiated 2017-12-18, terminated 2019-04-09. It names one FDA 510(k) clearance: K953239. Product: Modular E 170 Analyzer: Immunoassay Chemistry Analyzer, Chemistry (Photometric, Discrete) for clinical use, Catalog Numbers: 03739040692, 03023109001, 05023599001, 05023572001, 03739040001 & 03023109973.. Reason: Complaints have been received concerning questionable results on analyzers due to customers utilizing sample 13 mm tubes without tube adapters..
- Recalling firm
- Roche Diagnostics Corporation
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2017-12-18
- Terminated
- 2019-04-09
- Firm location
- Indianapolis, IN, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Modular E 170 Analyzer: Immunoassay Chemistry Analyzer, Chemistry (Photometric, Discrete) for clinical use, Catalog Numbers: 03739040692, 03023109001, 05023599001, 05023572001, 03739040001 & 03023109973.
Reason for recall
Complaints have been received concerning questionable results on analyzers due to customers utilizing sample 13 mm tubes without tube adapters.