Recall Z-1267-2021— Ortho-Clinical Diagnostics, Inc.
Class II · Terminated
Class II FDA medical device recall by Ortho-Clinical Diagnostics, Inc., initiated 2021-02-12, terminated 2022-01-24. FDA's recall record names no specific 510(k) clearance for this event. Product: VITROS SARS-CoV-2 Antigen Reagent Pack - The VITROS Immunodiagnostic Products SARS-CoV-2 Antigen Reagent Pack when used in combination with the VITROS Immunodiagnostic Products SARS-CoV-2 Antigen Calibrator is a chemiluminescent immunoassay intended for the qualitative detection of SARS-CoV-2 nucleocapsid protein antigens in nasopharyngeal (NP) specimens. For emergency authorization use only (EAU). Product Code: 6199941. OUS Product.. Reason: VITROS Immunodiagnostic Products SARS-CoV-2 Antigen False Reactive Results Associated When Using Remel M4RT Viral Transport Media.
- Recalling firm
- Ortho-Clinical Diagnostics, Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2021-02-12
- Terminated
- 2022-01-24
- Firm location
- Rochester, NY, United States
Product description
VITROS SARS-CoV-2 Antigen Reagent Pack - The VITROS Immunodiagnostic Products SARS-CoV-2 Antigen Reagent Pack when used in combination with the VITROS Immunodiagnostic Products SARS-CoV-2 Antigen Calibrator is a chemiluminescent immunoassay intended for the qualitative detection of SARS-CoV-2 nucleocapsid protein antigens in nasopharyngeal (NP) specimens. For emergency authorization use only (EAU). Product Code: 6199941. OUS Product.
Reason for recall
VITROS Immunodiagnostic Products SARS-CoV-2 Antigen False Reactive Results Associated When Using Remel M4RT Viral Transport Media
Data sourced from openFDA. This site is unofficial and independent of the FDA.