Recall Z-1267-2021— Ortho-Clinical Diagnostics, Inc.

Class II · Terminated

Class II FDA medical device recall by Ortho-Clinical Diagnostics, Inc., initiated 2021-02-12, terminated 2022-01-24. FDA's recall record names no specific 510(k) clearance for this event. Product: VITROS SARS-CoV-2 Antigen Reagent Pack - The VITROS Immunodiagnostic Products SARS-CoV-2 Antigen Reagent Pack when used in combination with the VITROS Immunodiagnostic Products SARS-CoV-2 Antigen Calibrator is a chemiluminescent immunoassay intended for the qualitative detection of SARS-CoV-2 nucleocapsid protein antigens in nasopharyngeal (NP) specimens. For emergency authorization use only (EAU). Product Code: 6199941. OUS Product.. Reason: VITROS Immunodiagnostic Products SARS-CoV-2 Antigen False Reactive Results Associated When Using Remel M4RT Viral Transport Media.

Recalling firm
Ortho-Clinical Diagnostics, Inc.
Classification
Class II
Status
Terminated
Initiated
2021-02-12
Terminated
2022-01-24
Firm location
Rochester, NY, United States

Product description

VITROS SARS-CoV-2 Antigen Reagent Pack - The VITROS Immunodiagnostic Products SARS-CoV-2 Antigen Reagent Pack when used in combination with the VITROS Immunodiagnostic Products SARS-CoV-2 Antigen Calibrator is a chemiluminescent immunoassay intended for the qualitative detection of SARS-CoV-2 nucleocapsid protein antigens in nasopharyngeal (NP) specimens. For emergency authorization use only (EAU). Product Code: 6199941. OUS Product.

Reason for recall

VITROS Immunodiagnostic Products SARS-CoV-2 Antigen False Reactive Results Associated When Using Remel M4RT Viral Transport Media