Recall Z-1265-2021— Philips North America Llc

Class II · Terminated

Class II FDA medical device recall by Philips North America Llc, initiated 2021-02-22, terminated 2023-02-09. It names one FDA 510(k) clearance: K180015. Product: Incisive CT Computed Tomography X-ray System Model: 728143. Reason: Actuators in the Incisive gantry could result in the gantry failure to self-lock if both of the 2 sides actuators fail. In some cases in systems with affected components, the gantry may tilt forward slowly, in a worst case, reaching to the tilt limit angle +31degrees.

Recalling firm
Philips North America Llc
Classification
Class II
Status
Terminated
Initiated
2021-02-22
Terminated
2023-02-09
Firm location
Cambridge, MA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Incisive CT Computed Tomography X-ray System Model: 728143

Reason for recall

Actuators in the Incisive gantry could result in the gantry failure to self-lock if both of the 2 sides actuators fail. In some cases in systems with affected components, the gantry may tilt forward slowly, in a worst case, reaching to the tilt limit angle +31degrees